    Validation of Haier BP3BG1-A automatic upper arm blood pressure monitor for clinical use and self-measurement[J]. Chinese Heart Journal, 2015, 27(1): 92-094.
    Citation: Validation of Haier BP3BG1-A automatic upper arm blood pressure monitor for clinical use and self-measurement[J]. Chinese Heart Journal, 2015, 27(1): 92-094.

    海尔BP3BG1-A全自动上臂电子血压计准确性的临床评价

    Validation of Haier BP3BG1-A automatic upper arm blood pressure monitor for clinical use and self-measurement

    • 摘要: 目的:按照《欧洲高血压协会国际质量控制方案2010修正版(ESH-IP 2010)》,评估海尔BP3BG1-A全自动上臂电子血压计临床应用的准确性。方法: 测量者采用水银血压计和受试血压计先后测量33名被试者同一左侧上臂收缩压和舒张压。根据ESH-IP 2010,每位被试者测量3对血压,共收集99例次受试血压计与水银血压计的测值。结果: 受试血压计测得73次(74%)收缩压与水银血压计测值差在5 mmHg以内,98次(99%)在10 mmHg以内,99次(100%)在15 mmHg以内;86次(87%)舒张压与水银血压计测值差5 mmHg以内,98次(99%)差10 mmHg。受试血压计测得收缩压与水银血压计测值差为(3.1±3.7) mmHg;舒张压测值差为(-0.9±3.7) mmHg。采用受试血压计与水银血压计的测值2~3次的差别都在5 mmHg以内的受试者数目分别为26名(79%)和29名(88%)。结论: 受试血压计对临床成人患者血压测量的准确性达到《 ESH-IP 2010》的验证标准。

       

      Abstract: AIM:To validate the Haier BP3BG1-A automatic upper arm blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. METHODS: Sequential same-left-arm measurements of systolic and diastolic blood pressure were obtained in 33 participants using the mercury sphygmomanometer and the test device. According to the validation protocol, 99 pairs of the test device and reference blood pressure measurements (three pairs for each of the 33 participants) were obtained in the study. RESULTS: The device produced 73, 98 and 99 measurements within 5, 10 and 15 mmHg for systolic blood pressure (SBP) and 86, 98 and 99 for diastolic blood pressure (DBP), respectively. The mean SD device-observer difference was (3.1±3.7) mmHg for SBP and (-0.9±3.7) mmHg for DBP. The number of subjects with two or three of the device-observer difference within 5 mmHg was 26 for SBP and 29 for DBP. CONCLUSION: Haier BP3BG1-A automatic upper arm blood pressure monitor can be recommended for clinical use and self-measurement in an adult population on the basis of the ESH-IP revision 2010.

       

    /

    返回文章
    返回